Laboratory Water Specs - LAB WATER SYSTEM
Lab Water Systems
Everything you need to make science grade water
An APSWATER Micro-Website™
A US Government Supplier DUNS 787700418

Rated "A+" with the BBB
33 years, Zero Complaints
Click here to see our record
Lab Water System ON SALE NOW
Introducing the Polaris Lab Water Systems
High Purity Water Made In The USA.
Click here for more info.

Laboratory Water Specs



ASTM Type I, II and III Water Specifications 2010

  Type I Type II Type III Type IV
Electrical conductivity, max, µS/cm 0.056 1.0 4.0 5.0
Electrical resistivity, min 18 1.0 0.25 0.2
pH * * * 5.0 to 8.0
Total organic carbon (TOC), max, µg/L 50 50 200 no limit
Sodium, max, µg/L 1 5 10 50
Chlorides, max, µg/L 1 5 10 50
Total silica, max, µg/L 3 3 500 no limit

Bacterial contamination

When bacteria levels need to be controlled, reagent grade types should be further classified as follows:

    Type A Type B Type C
Maximum heterotrophic bacteria count   10/1000 mL 10/100 mL 100/10 mL
Endotoxin, EU/mlB   0.03 0.25 not applicable


*pH is not a significant indicator when comparing ionic contamination.

Also See : 
Laboratory Water Specs

All Images are representative and may or may not be of actual product.

Tags : LAB WATER SYSTEM | LABORATORY | WATER | SPECS

LAB-WATER-SYSTEM.COM content (C)2009-2024 APS Water Services Corporation - All rights reserved
Website Software - Multi-Website Content Manager - Wilson Web Engineering (C)2012-2024

Micro-Website is a trademark of Wilson Web Engineering

laboratory water specs
lab-water-system.com micro-website™
4/23/2024
10:22:43 PM

laboratory water specs | lab-water-system.com
Google Apps

Barnstead Filters and Parts


Other APS Web Properties
Barnstead-Water.com lab-Water-filters.com lab-water System.com LabFilter.com Millipore-alternative.com Durastill Water Distillers Ahlstrom-Filters.com