ASTM Type I, II and III Water Specifications 2010

  Type I Type II Type III Type IV
Electrical conductivity, max, µS/cm 0.056 1.0 4.0 5.0
Electrical resistivity, min 18 1.0 0.25 0.2
pH * * * 5.0 to 8.0
Total organic carbon (TOC), max, µg/L 50 50 200 no limit
Sodium, max, µg/L 1 5 10 50
Chlorides, max, µg/L 1 5 10 50
Total silica, max, µg/L 3 3 500 no limit

Bacterial contamination

When bacteria levels need to be controlled, reagent grade types should be further classified as follows:

    Type A Type B Type C
Maximum heterotrophic bacteria count   10/1000 mL 10/100 mL 100/10 mL
Endotoxin, EU/mlB   0.03 0.25 not applicable


*pH is not a significant indicator when comparing ionic contamination.

Also See : 
Laboratory Water Specs

All Images are representative and may or may not be of actual product.

Tags : LAB WATER SYSTEM | LABORATORY | WATER | SPECS

LAB-WATER-SYSTEM.COM content (C)2009-2024 APS Water Services Corporation - All rights reserved
Website Software - Multi-Website Content Manager - Wilson Web Engineering (C)2012-2024

Micro-Website is a trademark of Wilson Web Engineering

laboratory water specs
lab-water-system.com micro-website™
12/13/2024
10:06:25 PM

laboratory water specs | lab-water-system.com
Google Apps

Barnstead Filters and Parts


Other APS Web Properties
Barnstead-Water.com lab-Water-filters.com lab-water System.com LabFilter.com Millipore-alternative.com Durastill Water Distillers Ahlstrom-Filters.com